A failed PCR is usually a small number of causes: the template, the primers, the components, the cycling or the readout. Work through them in a fixed order so you do not change three things at once.
Check the template first
Is the template present, intact and in the expected concentration? A nanodrop reading can be misleading for dirty samples. Run a small amount on a gel or use a quantitative method you trust.
Is the template the right one? A tube label is not evidence. If the expected product is absent and a different product appears, consider a sample mix-up or contamination. Include a no-template control and, when relevant, a positive control.
Check the primers
Primers degrade with repeated freeze–thaw cycles. Check the concentration and the storage. If a primer pair worked before and now fails, try a fresh aliquot.
A primer that is specific in silico can still fail because of a mismatch in the actual sample, a secondary structure or a wrong annealing temperature. Review the design and, if possible, test a second pair.
Check the components
Use the same master mix that worked before. A new lot, a thawed enzyme or a missing component can cause failure. Prepare a master mix for the number of reactions plus one, and keep the enzyme cold.
Magnesium concentration, dNTP quality and pH can all matter. If the protocol is new, follow the supplier's recommendation first, then optimise one variable at a time.
Check the cycling
The annealing temperature is the most common variable. A gradient PCR can identify a better temperature in one run. Extension time depends on product length and polymerase speed.
Denaturation should be long enough to separate the template but not so long that the enzyme is damaged. A high GC template may need a higher denaturation temperature or an additive.
Check the readout
A gel can fail for reasons unrelated to PCR: the gel was run too long, the loading dye was missing, the ladder was wrong or the DNA was below the detection limit. A qPCR amplification curve can be flat because of a threshold setting, not because the reaction failed.
Always run a positive control and a ladder. If the positive control works and the sample does not, the problem is in the sample. If the positive control fails, the problem is in the reaction or the readout.
A troubleshooting order
- Positive control, negative control and ladder.
- Template quality and quantity.
- Primer integrity and concentration.
- Master mix and enzyme.
- Annealing temperature and extension time.
- Additives for GC-rich templates.
- Readout and gel conditions.
Change one variable per experiment and record the result. A troubleshooting log is more useful than a collection of failed tubes.
A symptom-first view
The order above is a checklist. In practice, the symptom narrows the search.
No band at all with a working positive control points to the sample: wrong template, inhibitors, degraded DNA or a sample mix-up. No band and a failed positive control points to the reaction: a missing component, a dead enzyme or a cycling error.
A band of the wrong size points to specificity: a misprimed product, a pseudogene, genomic DNA in a cDNA reaction or a primer that matches a paralogue. A smear points to too little stringency, too much template, degraded template or too many cycles.
A band in the no-template control is contamination. Stop and find the source before running more samples. A band that appears only in the negative control and nowhere else can also be a primer dimer that is visible on the gel.
Contamination versus failure
Contamination and failure look similar on a gel but need opposite responses. If the negative control is clean and the sample fails, do not assume contamination. If the negative control has a band, do not assume the sample is fine because its band looks brighter.
Keep separate areas and pipettes for pre- and post-amplification work when the volume justifies it. Use filter tips, aliquot reagents and open tubes one at a time. A contamination event is easier to prevent than to trace.
Recording an optimisation
An optimisation is an experiment, so record it like one. Note the template concentration, primer pair and concentration, master mix lot, annealing temperature, extension time, cycle number and the result for each condition.
Change one factor at a time, or use a small design that lets you separate the effects. A gradient PCR changes temperature across a block, which is efficient, but it does not test primer concentration at the same time. If you change two factors together and the reaction improves, you do not know which one helped.
When the product is the wrong size
First check the expected size against the primers and the reference sequence. A calculation error is common. Then check whether the template contains a paralogue, a pseudogene or an alternative splice form that the primers can amplify.
If the extra band is faint, increasing the annealing temperature or reducing cycle number may be enough. If it is strong, redesign the primers or choose a region with better specificity. Gel extraction of the correct band is possible, but it does not fix a primer that amplifies the wrong target efficiently.
Key points
- Use positive and negative controls in every run.
- Check template, primers, components, cycling and readout in that order.
- Change one variable at a time.
- A gel failure is not always a PCR failure.
- Record the conditions and the outcome.
Common mistakes
- Changing the annealing temperature and the primer pair at the same time.
- Using a degraded or contaminated template.
- Forgetting the no-template control.
- Assuming a nanodrop reading means the template is good.
- Blaming the polymerase before checking the readout.
When to ask for help
If a control fails, if the product is the wrong size, or if you need to optimise a difficult template, ask for a review. You can order an analysis or read the primer design guide. The free PCR master mix calculator helps with volumes, and the PCR troubleshooting article covers primer-specific issues.