How the tool reads a gel image
The picture is converted to one brightness value per pixel, then the region you outline is split into equal-width lanes. Averaging each lane across its own columns gives one value per row: the lane profile, which is what a densitometer trace looks like turned on its side, with a band showing as a peak.
Bands darker than their background are flipped automatically, which the tool decides by comparing the mean brightness with the median: a gel full of dark bands has a mean well below the background level that the median reports. A straight baseline is then drawn between the 5th percentile of the first tenth of the profile and the 5th percentile of the last tenth, because the top and bottom of a lane are where gel background sits, and everything is measured above that line. Peaks are the local maxima that reach 5 % of the brightest band, and each peak is integrated between the minima on either side of it.
Why the answer is relative, not absolute
A band area is brightness-times-rows, which has no physical unit: it depends on the exposure, the stain, the camera and the thickness of the gel. The defensible statement is always comparative, so the tool reports every lane as a multiple of a control lane you choose, which is the same logic as the delta-delta-Ct calculation for qPCR.
A loading control does a different job. If you outline the stripped and re-probed housekeeping band as a second rectangle, each lane is first divided by its own loading area before being compared with the control, which cancels how much sample went into that lane while leaving the biology. It does not correct transfer efficiency, uneven staining across the gel, or a band that has run out of the linear range of detection.
Getting a number you can defend
Draw the region over the bands only, from the top of the gel to below the lowest band you want counted, and keep it clear of the lanes on either side: equal-width lanes are an assumption, and a rectangle that includes the well or the dye front will bend the baseline. If the brightest pixel of a band is pure white, the band is saturated and its reported area is a floor rather than a measurement — shorten the exposure and quantify that instead.
A phone photograph of a gel carries its lighting gradient inside the baseline the tool tries to fit, so for anything that leaves the lab use the imager’s own software and keep the raw file. This page is for the quick honest look at a gel you already have: is the knockdown as strong as it looks, did the second PCR lane actually carry more product, and how much does the answer move if you change the baseline.