What is reported
Enter a peptide in one-letter code. The calculator sums residue and water masses from the shared proteomics tables and reports the monoisotopic and average neutral masses, the singly, doubly and triply protonated m/z values, and the elemental composition.
Monoisotopic mass uses the lightest stable isotope of each element; average mass uses natural abundances. High-resolution instruments report monoisotopic peaks, while low-resolution work often uses average mass.
Modifications
N-terminal acetylation (+42.010565 Da), C-terminal amidation (−0.984016 Da), carbamidomethyl cysteine (+57.021464 per C), oxidised methionine (+15.994915 per M) and phosphorylation (+79.966331) at the S/T/Y positions you list are applied to the neutral mass and to the composition. Positions are 1-based; a position that is not S, T or Y is reported and ignored rather than silently modified.
The masses come from the same constants used by the Proteomics app, and m/z is (neutral mass + z × proton mass) / z with the proton mass 1.00727646688, so [M+H]+ is the neutral mass plus one proton, not plus a hydrogen atom.
Composition
The elemental formula includes modification atoms and reports phosphorus separately. The hydrogen count is the neutral molecule’s count; the ionic forms add the charge-carrying protons shown in the m/z column.