What qPCR efficiency means
In an ideal reaction the amount of product doubles every cycle, an efficiency of 100%. To measure it, amplify a dilution series of known template amounts and plot Ct (cycle threshold) against the log₁₀ of the starting quantity. The points fall on a line, and its slope gives the efficiency: E = 10^(−1/slope) − 1.
A slope of −3.32 corresponds to 100% efficiency. A slope of −3.1 is about 110% and −3.6 about 90%.
Acceptable ranges
Most guidelines (including MIQE) accept 90–110% efficiency with R² of at least 0.98, preferably 0.99. Efficiency above 110% usually points to pipetting error, inhibitors at the high-concentration points or a poorly set baseline rather than real amplification above doubling. Low efficiency suggests poor primer design, inhibitors or a suboptimal annealing temperature.
Using the result
The ΔΔCt method assumes both the target and reference gene amplify at about 100% efficiency. If their efficiencies differ, use an efficiency-corrected model such as Pfaffl. Enter at least three dilution points; five or more across several orders of magnitude gives a reliable estimate. Use the mean of technical replicates for each dilution.