RPM versus RCF
Protocols should give centrifuge speed as relative centrifugal force (RCF, in ×g) because the force on a sample depends on the rotor radius as well as the speed. The same RPM gives very different forces in a small benchtop rotor and a large floor-standing one.
The formula
RCF = 1.118 × 10⁻⁵ × r × RPM², where r is the radius from the rotor axis to the sample in centimetres. Inverting it gives RPM = √(RCF / (1.118 × 10⁻⁵ × r)). For a radius of 8.4 cm, 10,000 RPM is about 9,400 ×g.
Which radius?
The maximum radius (the tip of the tube, r max) is most often quoted, but the average radius is also used. Check the rotor’s data sheet and use the same value as the protocol author.