ELISA (enzyme-linked immunosorbent assay) measures a protein, antibody or hormone in a sample by capturing it with antibodies on a plastic plate and turning that binding into a colour you can read. It is one of the most widely used assays in diagnostics, vaccine research and biotech quality control, and one of the first immunoassays students learn. This guide covers the principle, the four main types, a general sandwich ELISA protocol, how to analyse the standard curve and how to troubleshoot the usual problems.
The principle of ELISA
Every ELISA has three ideas in common:
- Something is immobilised on the wells of a 96-well polystyrene plate: an antigen or a capture antibody.
- Specific binding between an antibody and its antigen holds the target in place while everything else is washed away.
- An enzyme attached to a detection antibody, most often horseradish peroxidase (HRP) or alkaline phosphatase (AP), converts a colourless substrate into a coloured product. The colour is proportional to the amount of target.
With HRP and TMB substrate the wells turn blue, then yellow when the reaction is stopped with acid, and the plate is read at 450 nm.
The four types of ELISA
| Type | What is coated | Detects | Typical use |
|---|---|---|---|
| Direct | Antigen | Antigen, with an enzyme-labelled primary antibody | Quick screening; fewer steps, lower sensitivity |
| Indirect | Antigen | Antibodies in a sample, with an enzyme-labelled secondary | Measuring antibody responses (serology, vaccine studies) |
| Sandwich | Capture antibody | Antigen between two antibodies | Cytokines, biomarkers; most sensitive and specific |
| Competitive | Antigen or antibody | Small molecules and antigens with one epitope | Hormones, drugs, toxins; signal falls as target rises |
The sandwich format needs two antibodies that bind different epitopes on the target, which is why commercial "matched pairs" exist.
A general sandwich ELISA protocol
Always follow the kit or antibody datasheet; this is the usual order of steps.
- Coat the plate with capture antibody (typically 1–4 µg/mL in carbonate or PBS buffer), 100 µL per well, overnight at 4 °C.
- Wash three times with PBS containing 0.05% Tween-20 (PBST).
- Block with 1–5% BSA or another blocking buffer for 1–2 hours to cover remaining binding sites.
- Add standards and samples in duplicate or triplicate, 100 µL per well, 2 hours at room temperature. Include blanks.
- Wash, then add the biotinylated detection antibody for 1–2 hours.
- Wash, then add streptavidin–HRP for 20–30 minutes, protected from light.
- Wash thoroughly; residual HRP is the commonest cause of high background.
- Add TMB substrate and incubate 10–30 minutes until the top standard is clearly blue.
- Stop with 1 M sulfuric or 2 M phosphoric acid and read at 450 nm within 30 minutes, with 570 nm or 630 nm as a reference wavelength if your reader supports it.
Analysing ELISA data: the standard curve
- Average the replicate absorbances and subtract the blank.
- Plot the standards: concentration on the x-axis (usually log scale) and absorbance on the y-axis.
- Fit a four-parameter logistic (4PL) curve. A straight line only fits the middle of the range and gives wrong results at the ends.
- Read each sample's concentration from the curve and multiply by its dilution factor.
- Only report samples that fall inside the range of the standards. Dilute and re-run samples above the top standard.
Check that replicate wells agree, with a coefficient of variation (CV) under about 10–15%. Our free Standard Curve Calculator fits curves in the browser if you want to try it with your own numbers.
Troubleshooting ELISA
| Problem | Likely causes | What to try |
|---|---|---|
| High background in every well | Too little washing, blocking too weak, detection reagent too concentrated | Wash more thoroughly, block longer, titrate the detection antibody |
| No signal, including standards | A missed reagent, expired substrate, sodium azide in buffers (it inhibits HRP) | Check each step, use fresh TMB, remove azide |
| Weak signal | Short incubations, over-diluted antibodies, standards degraded | Extend incubations, prepare fresh standards |
| Poor replicates (high CV) | Pipetting errors, bubbles, wells drying out | Use calibrated pipettes, change tips, never let wells dry |
| Edge effects | Uneven temperature across the plate | Seal plates and incubate away from vents; avoid stacking |
ELISA vs Western blot
ELISA is quantitative and handles many samples at once, which suits measuring a known protein in serum or culture supernatant. A Western blot separates proteins by size first, so it confirms identity and shows modified or degraded forms, but it is only semi-quantitative. Labs commonly use both.
Learn ELISA hands-on
Reading about ELISA is not the same as running a clean plate. Our ELISA & Immunoassay Techniques course is a short, hands-on training with a certificate: you coat, block and develop your own plates, fit a 4PL curve and troubleshoot real results. For the protein work around it, see Protein Expression, Purification & Western Blotting.
Written by the Solagene team for research and education. Found an error? Tell us and we will correct it.