Gating in flow cytometry is the process of selecting the events that belong to the population you want to analyse. A gate is a decision rule, and every downstream percentage or fluorescence measurement depends on it.
The order of gates
Most analyses start with the same broad sequence. First remove debris and very small events using forward scatter (FSC) and side scatter (SSC). Then remove doublets using a plot of FSC-area against FSC-height or FSC-width. Then select live cells with a viability dye. Then select the cell type of interest with surface or intracellular markers.
The order matters because a later gate can only see the events that passed the earlier ones. If debris is included in the singlet gate, the singlet percentage is wrong. If dead cells are included in the live gate, the marker percentages are distorted.
Start with the physics of the instrument, then the biology. Scatter gates are usually drawn first because they do not depend on fluorescence compensation.
Scatter gates
FSC roughly reflects cell size, and SSC reflects internal complexity or granularity. Lymphocytes, monocytes and granulocytes often form separate clusters in a blood sample. In cultured cells, the clusters may be less distinct.
Draw the scatter gate generously enough to include the population of interest and tightly enough to exclude debris and obvious aggregates. Check the gate on a sample with and without the treatment, because a treatment can change cell size or granularity. A gate that is too tight can make a real population disappear.
Doublet removal
Two cells passing the laser together produce one event with roughly twice the signal. Doublets can look like a highly fluorescent population and can distort both percentages and sorting.
A singlet gate uses the fact that a doublet has a different relationship between pulse area and pulse height or width than a single cell. Plot area against height and select the diagonal population. The exact axis names depend on the instrument.
Remove doublets before analysing fluorescence. A doublet gate placed after a marker gate cannot recover the single cells that were lost.
Viability
Dead cells bind antibodies nonspecifically and release fluorescence that can contaminate neighbouring channels. A viability dye that cannot enter live cells labels dead cells, which are then excluded.
The viability gate should be placed before the marker gates. If the dye is added after fixation, check that the dye works in fixed cells. Some dyes are live-cell reagents only.
A viability gate is not a substitute for good sample preparation. A sample with many dead cells will have few live events, and the remaining population may not represent the original sample.
Fluorescence gates and controls
Fluorescence gates need controls. An unstained sample shows the background autofluorescence and the spread of the negative population. A single-stained sample for each fluorophore defines the positive population and helps set the compensation.
For a marker that is expressed continuously rather than as a clear positive and negative population, a gate based on a negative control or an isotype control is common. Isotype controls control for nonspecific binding, but they do not control for the biology of the marker. Use them with care.
Fluorescence-minus-one (FMO) controls are often more informative for setting a gate in a multicolour panel. An FMO sample contains every fluorophore except the one being gated, so it shows the spread caused by the other channels.
Documenting a gating strategy
A gating strategy should be reproducible. Record the order of the gates, the axes, the gate boundaries and the number of events in each gate. For a publication, show the gating tree in a supplementary figure.
Use the same gate boundaries across samples in a comparison when the biology allows it. If a gate must be adjusted for a sample, say why and show the change. Moving a gate to make a population appear is not analysis.
Report the number of events analysed. A percentage based on a small number of events is uncertain. For rare populations, collect enough events to make the estimate meaningful and state the limit of detection.
Common gating mistakes
- Drawing a gate around a cluster without a biological reason.
- Including debris or doublets in the analysis.
- Setting a positive gate from an unstained sample alone.
- Using an isotype control as if it were a biological negative.
- Changing gates between groups without reporting it.
A practical workflow
- Plot FSC against SSC and exclude debris.
- Remove doublets with an area-versus-height plot.
- Exclude dead cells with a viability dye.
- Gate the cell type of interest with a marker and a control.
- Check the gate with an FMO or a biological control.
- Record the gate hierarchy and event counts.
- Export the statistics and the gating tree.
Reporting percentages
A percentage is only meaningful with the denominator. Report the parent population, the number of events in the parent gate and the number in the child gate. A statement that "40 percent of cells were positive" is incomplete if the reader cannot tell whether that means 40 percent of all events, of live cells or of a specific subset.
For rare populations, report both the percentage and the event count. A value of 1 percent based on 50 events is much less certain than 1 percent based on 5,000 events. State the acquisition target and whether it was met.
Key points
- Gate in a deliberate order: scatter, singlets, viability, markers.
- Doublets and dead cells distort fluorescence measurements.
- Fluorescence gates need appropriate controls.
- Keep gates consistent and document any change.
- Report event counts, not only percentages.
Common mistakes
- Gating on fluorescence before removing doublets.
- Using an unstained control to set a positive gate.
- Treating an isotype control as a biological negative.
- Hiding a small event count behind a percentage.
- Adjusting gates until the expected result appears.
When to ask for help
If a population is not where you expect it, if the positive and negative populations overlap, or if the panel needs a new fluorophore, ask for a review. You can order an analysis or read the compensation guide. The Flow Cytometry studio helps plot and gate FCS files, and the cell counting guide covers the sample preparation.