Before you start
- An Applied Biosystems .ab1 or .abi trace file.
- A reference FASTA if you plan to compare the read.
- The opposite-strand read or a second trace if you need to confirm a difference.
Step by step
Open Trace and load the file
Drop an .ab1 file or press Load demo (400-base synthetic trace). The app reports an error if the file is not ABIF or contains no base calls.
Inspect the electropherogram
The trace shows base letters, quality bars and the four dye channels. Scroll through the whole read; the start and end are usually the weakest regions.
Read mixed peaks
The app marks mixed peaks with IUPAC codes and a secondary base. A mixed peak can be a heterozygous variant, a mixed template or a sequencing artefact.
Open Trim & export
Choose Mott or window quality trimming. The app exports trimmed and full FASTA, plus FASTQ with the instrument qualities, so the trimming is reproducible.
Open Compare and load a reference
Paste a reference sequence or load a FASTA. The app aligns the read in both orientations and reports a difference table with reference position, read position, quality and confidence.
Confirm every difference
A difference is a candidate, not a validated variant. Check the trace at that position and, where possible, the opposite strand before reporting it.
How to read the results
| Output | What it tells you |
|---|---|
| Electropherogram | The raw signal and the called bases. A clean single peak is strong evidence; a small shoulder under a peak is not. |
| Mixed-peak table (Position, Called, Secondary, IUPAC) | Positions where more than one base was called. The IUPAC code tells you which bases are present, not which one is biological. |
| Trim summary | How many bases were removed and why. Keep the full-length export so the trimming can be checked. |
| Difference table (Ref position, Reference, Read, Type, Quality, Confidence) | Where the read differs from the reference. Low quality or a mixed peak at the same position weakens the call. |
Common problems
- The file is not ABIF.
- The app accepts .ab1 or .abi traces from Applied Biosystems. Convert the file with the instrument software; a FASTA or PDF trace is not a trace file.
- No base calls are found.
- The trace may contain raw channels without called bases. Re-export it with base calling enabled.
- The quality length does not match the bases.
- The file is internally inconsistent or was edited. Re-export the original trace rather than patching the arrays.
- Compare refuses without a reference.
- Paste a reference sequence or load a reference FASTA. The comparison needs something to align to.
- A difference looks real but is only in the read.
- Check the trace quality and the opposite strand. The local alignment is not a validated variant caller.
Limits
- Base calls and qualities are the instrument software’s; the viewer does not call bases.
- Local alignment is not a validated variant caller.
- Confirm every difference on the trace or the opposite strand.
The app records its tool, version, parameters and input checksum in the methods log. The documentation on reproducibility explains how that record becomes a methods paragraph, and the Report Builder collects the figures into one export.
How to cite this analysis
Cite the Workbench using the Zenodo DOI on the Cite page (/cite). Report the instrument, base-calling software, trimming method and the reference accession. Citation, DOI and BibTeX.
Related reading
- FASTQ Read Processing: a step-by-step guide
- Sequence Analysis: a step-by-step guide
- Variant Analysis: a step-by-step guide
- Peptide Analysis: a step-by-step guide
Last reviewed: 2026-11-10. If a screen does not match this guide, the app is the source of truth; tell us so the guide can be corrected.