How a count becomes cells per mL
A counting chamber is a fixed volume of liquid held between a glass slide and a coverspring of a known height. The standard Improved Neubauer is 0.1 mm deep and each of the four large corner squares measures 1 mm², so one square holds 0.1 mm³ — 100 nanolitres, or one ten-thousandth of a millilitre.
That single fact is where the familiar ×10⁴ comes from: take the mean count per square, multiply by your dilution factor and by 10,000, and you have cells per mL. Averaging over several squares matters more than the number of squares suggests, because cells settle and clump; the coefficient of variation between your squares is shown so you can see whether the suspension was mixed well enough to trust the mean.
Trypan blue and what viability really says
Dye exclusion works on membrane integrity: a cell with a sound plasma membrane keeps 0.4 % trypan blue out and stays bright, while a dead or badly damaged one turns blue. Counting both in the same squares gives viability as live ÷ (live + dead).
It is worth being precise about what that percentage is not. It is not a measure of whether the cell can still divide — a stressed cell can form a colony while staining blue, and a cell with a leaky membrane may still be counted clear. For most mammalian work the number you seed with is the viable concentration, because that is the one that predicts how many cells attach and grow; the total concentration tells you how much debris and how many dead cells are in the tube, which matters for downstream purification and for the day-to-day quality of a line.
From the count to a seeding plan
Volume to seed = cells wanted ÷ viable concentration, so 2×10⁵ cells from a 1×10⁶ per mL suspension is 0.2 mL = 200 µL. The tool does this for you and stops when the arithmetic will not fit your vessel: if the volume needed is more than the well holds, you cannot dilute your way out — spin the cells down and resuspend in a smaller volume, or split the seed across several wells.
The same concentration also gives the diluent to add to reach a target density (C1V1 = C2V2): 10 mL at 1×10⁶ per mL diluted to 2×10⁵ per mL needs 40 mL of medium, for a final 50 mL. Enter the chamber depth and square area if you use something else — a Fuchs–Rosenthal chamber is 0.2 mm deep, so it reports half the concentration for the same counts, and the small central squares used for red cells are 0.0025 mm².