The standard curve
Absolute quantification uses standards of known quantity (plasmid copies, synthetic RNA or a purified amplicon) run alongside the unknowns. A linear regression of Ct against log₁₀(quantity) gives Ct = intercept + slope × log₁₀(quantity). Replicates are averaged at each standard level before the fit, so a triplicate level counts once rather than three times.
The slope predicts amplification efficiency: E = 10^(−1/slope) − 1, reported as a percentage. A slope of −3.322 corresponds to 100% efficiency, where the product doubles every cycle. The page warns when efficiency falls outside 90–110% or R² is below 0.98.
Reading unknowns and scaling
For an unknown Ct, the fitted line is inverted: quantity = 10^((Ct − intercept) / slope). The result is flagged as extrapolated when its Ct lies outside the range covered by the standards, because the line beyond that range is not supported by measurements.
If you used only part of an elution in the reaction, enter the elution volume and the volume used per reaction. The calculator then reports copies per reaction multiplied by elution volume ÷ reaction volume, giving the amount in the whole elution. An optional template length converts copy numbers to nanograms of double-stranded DNA using 660 g/mol per base pair.
Limits and QC
The copy number is only as good as the standards: their identity, purity, dilution accuracy and the assumption that the target amplifies like the standard. Inhibitors, a baseline set too high or a primer dimer can distort the curve without an obvious warning. Always check the melt curve and the standard curve together.
This is research-grade arithmetic, not a diagnostic or clinical measurement. The page does not know your instrument, chemistry or sample matrix.