Quantifying a western blot means turning band intensities into a comparison between samples. The measurement is only as good as the linear range of the detection, the normalisation strategy and the replicates behind it.
What the band intensity represents
A western blot separates proteins by size, transfers them to a membrane and detects a specific protein with an antibody. The band intensity depends on the amount of protein, the affinity of the antibody, the transfer efficiency and the detection chemistry.
It is not a direct count of molecules. It is a signal that must be compared within the same membrane and the same exposure. Comparing a band from one membrane with a band from another is unreliable unless a common standard was run on both.
Stay inside the linear range
Film and digital detectors both saturate. A saturated band cannot be quantified because a stronger signal and a much stronger signal look the same. A very faint band is dominated by background and noise.
The practical approach is to run a dilution series of a control sample and detect it under the same conditions. The signal should increase with the amount of protein and then flatten. Use an exposure and a loading amount that sit in the linear part of that curve.
If the target changes a lot between conditions, the same exposure may be in range for one sample and saturated for another. Either dilute the samples or take multiple exposures and use the one that is in range for each comparison. Do not compare a saturated band with an unsaturated one.
Normalisation: housekeeping or total protein
Normalisation adjusts for differences in the amount of protein loaded. The traditional method uses a housekeeping protein such as actin, tubulin or GAPDH. The assumption is that the housekeeping protein does not change between conditions.
That assumption can fail. Housekeeping proteins change with treatment, cell type, confluence and stress. A treatment that affects the cytoskeleton can change actin; a metabolic treatment can change GAPDH. If the housekeeping protein changes, normalising to it can create or hide a difference.
Total protein staining is an alternative. It measures the overall protein in each lane and is less dependent on a single gene. Stain the membrane or a duplicate blot, quantify the total signal in each lane, and use that as the denominator. The method has its own biases, but it avoids the assumption that one gene is stable.
Whichever method you use, show the normalisation control in the figure or the supplement. A reader should be able to see that the loading was comparable.
Replicates and statistics
A western blot usually has technical replicates (the same sample run in several lanes) and biological replicates (independent samples or experiments). Technical replicates measure pipetting and detection precision. Biological replicates measure the variability of the biology.
Treating lanes as independent biological replicates inflates the sample size. If three lanes come from one lysate, they are not three independent experiments. The statistics should use the biological replicate as the unit.
With a small number of biological replicates, report the individual values and a measure of variability rather than relying on a significance test. A western blot with three biological replicates is common, but the uncertainty is large and the result should be described cautiously.
Background and specificity
A clean band at the expected size is necessary but not sufficient. A band can be the target, a modification, a degradation product or a cross-reacting protein. A second antibody or a knockout control can help establish specificity.
Subtract the local background around each band rather than a single background value for the whole membrane. Background can vary across the membrane. Use the same method for every band in the comparison.
If the membrane has stripes, speckles or a strong uniform background, the quantification may be dominated by artefact. Fix the detection problem before measuring.
Software and reporting
Imaging software can measure band intensity, but the settings matter. Use the same lane boundaries, background method and normalisation for every band. Save the raw image and the analysis file.
Report the antibody, dilution, catalogue number and supplier; the lysis buffer; the amount loaded; the transfer conditions; the detection method; the exposure; the normalisation method; the number of biological replicates; and the analysis software.
For a figure, show a representative blot and a quantification with individual points where possible. State whether the blot shown is representative and how many independent experiments were performed. Do not crop a blot in a way that hides a neighbouring band that would change the interpretation.
A practical workflow
- Lyse and quantify protein with a method suited to the buffer.
- Load equal amounts and include a normalisation control.
- Run the gel and transfer with a check of transfer efficiency.
- Block, incubate with primary and secondary antibodies, and detect.
- Take a series of exposures and choose one in the linear range.
- Quantify with consistent settings and subtract local background.
- Normalise and analyse biological replicates.
- Report the full method and the raw data.
Key points
- Quantify only within the linear range of detection.
- Normalise with a control whose stability you have checked.
- Treat biological replicates as the unit of statistics.
- Subtract local background consistently.
- Report the antibody, loading and analysis settings.
Common mistakes
- Quantifying a saturated band.
- Comparing bands from different membranes or exposures.
- Using a housekeeping protein that changes with treatment.
- Counting lanes as biological replicates.
- Cropping a blot to hide a neighbouring band.
When to ask for help
If a band is saturated, if the normalisation control changes, or if the quantification does not match the visual impression, ask for a review. You can order an analysis or read the western blot protocol guide. The free SDS-PAGE loading calculator and protein concentration calculator help with the upstream steps, and the gel band quantifier gives a quick intensity estimate.
Sources and further reading
Written by the Solagene team for research and education. Found an error? Tell us and we will correct it.